eca6879c4b8e777212f2118a97593ef3.jpg


The VHP head cover fumigation transfer window, a commonly used sterilization equipment in P3 and P4 workshops, is primarily designed for the transfer of head covers and protective clothing between low and high contamination probability areas after use. The dual doors feature interlocking mechanisms and inflatable seals, while the equipment is connected to the wall using a biological sealing method. With its VHP sterilization function, it effectively reduces the risk of biological contamination during the transfer of head covers.


d91af6d4afcbc2b05d6e82b49fe2f522.png




performance metrics

1. Sterilization cycle: The total process time for the preparation stage, sterilization stage, and residual removal stage should not exceed 120 minutes;

2. Sterilization level: sterilization value >6LOG (Bacillus stearothermophilus ATCC 12980);

3. Sterilization capacity: Capable of sterilizing up to 8 sets of positive pressure protective headgear simultaneously in a single operation;


Main Configuration

1. Power supply: AC 220V/50Hz, power 4kW;

2. Air source: 0.6MPa~0.8MPa clean compressed air;

3. Material: The internal chamber is made of high-quality 316L stainless steel, while the external main frame is made of high-quality 304 stainless steel;

4. Control system: SIEMENS brand PLC module, equipped with audit tracking, electronic signature, three-level permission management, and other functions;

5. Concentration sensor: It can detect the concentration inside the chamber during sterilization in real time, with a detectable concentration range below 1ppm. The brand is either VAISALA imported from Finland or ATI imported from the United States;

6. High-efficiency filter: The inlet and outlet duct ports are equipped with H14 grade high-efficiency filters;

7. Airtight door: An inflatable sealing door with dual protection using an electromagnetic lock and an inflatable sealing strip. The sealing strip is made of silicone material, with a usage frequency of >10,000 times and an inflation/deflation time of <5 seconds;

8. Fan: EBM brand centrifugal fan is adopted, with pressure and air volume meeting the equipment usage requirements.